Purpose:
To evaluate the relative partial pressure in oxygen (pO
2
) at the corneal surface under Boston XO
2
scleral lenses (SL) fitted with targeted clearances of 200 and 400 μm (SL200 and SL400).
Methods:
During this prospective study, the right eyes of eight normal subjects were fitted with SL200 and SL400. Clearance, validated after 5 minutes of wear with an optical coherence tomograph, was used with lens thicknesses to calculate transmissibility and estimate pO
2
. Corneal pO
2
s were measured with an oxygen electrode after 5 minutes of (1) corneal exposure to calibrating gases with various pO
2
or of (2) SL wear. Decays in pO
2
were modeled to an exponential. Linear regression between exponent k of these decays and calibrating gas pO
2
s allowed for the calculation of corneal pO
2
under SL. Differences between pO
2
s beneath SL200 and SL400 were tested with a mixed ANOVA.
Results:
The estimated transmissibility based on thicknesses and clearances (239.7 ± 34.7; 434.5 ± 33.2 μm) predicted a corneal pO
2
of 8.52 ± 0.51 and 6.37 ± 0.28% for SL200 and SL400. These values were close to measured pO
2
: 9.07 ± 0.86 and 6.19 ± 0.87% (mean ± SEM) (
P
<. 05) for SL200 and SL400, respectively. Both pO
2
measurements fall short of the theoretical values needed to prevent hypoxia during lens wear (9.9% and above).
Conclusions:
As shown
in vivo
for the first time, an 18‐mm scleral lens fitted with a 400‐μm clearance reduces the oxygen tension available to the cornea by 30% compared to a similar lens fitted with a 200‐μm clearance after 5 minutes of wear.