GVAX immunotherapy for prostate cancer is comprised of two genetically modified prostate cancer cell lines, CG1940 and CG8711, engineered to secrete granulocyte macrophage-colony-stimulating factor. As part of the matrix of potency assays, CG1940 and CG8711 are tested for the expression level of cell surface HER-2/neu using a quantitative flow cytometer assay. This assay reports the antibody binding capacity value of the cells as a measure of HER-2/neu expression using cells immediately after thawing from cryogenic storage. With optimized cell handling and staining procedure and appropriate system suitability controls, the assay was validated as a quantitative assay. The validation results showed that assay accuracy, specificity, precision, linearity, and range were suitable for the intended use of ensuring lot-to-lot consistency of HER-2/neu expression. Assay robustness was demonstrated using design of experiments that evaluated critical assay parameters. Finally, the assay was successfully transferred to a current good manufacturing practice Quality Control laboratory in a separate facility. Since the overall precision of this assay is better than that of ELISA methods and it can be performed with ease and high throughput, quantitative flow cytometer-based assays may be an appropriate immunological assay platform for Quality Control laboratories for characterization and release of cell-based therapies.