Objective: To establish a HPLC method for the determination of two potential genotoxic impurities Me toluenesulfonate and Et toluenesulfonate in ganciclovir condensates.Methods: Agilent Eclipse Plus C18 column (250 mm × 4.6 mm, 5 μm) was used with 0.1% phosphoric acid aqueous solution as mobile phase A and acetonitrile as mobile phase B.Diode array detector was used to detect the two targets at the detection wavelength of 225 nm.Results: There was a linear relationship between the peak areas of the two kinds of target substances in a certain range of mass concentration, and the detection limits (3 S/N) were 15.852 and 15.848 ng·mL-1, resp.The recoveries of actual samples were 101.3% ∼ 106.0% and 103.1% ∼ 106.7%, resp.Conclusion: This method is suitable for the determination of Me p-methylbenzenesulfonic acid and Et p-methylbenzenesulfonic acid in ganciclovir condensates.