The performance of enzymic method for serum sodium ion determination was evaluated.The accuracy, precision, off cap stability, linearity, correlation with ion selective electrode method, recovery rate and anti-interference were evaluated by the reagent of enzymic method for sodium ion determinationThe coefficient of variation (CV) was < 2%.Calibration period reached 5 d.Accuracy met Randox's quality control.Linearity scope was 100-160 mmol/L.Coefficient of correlation (r) was 0.9899, and enzymic method was correlated well with ion selective electrode method.The recovery rate was 100.9%.No interference appeared when bilirubin < 600 μmol/L, triglyceride < 20 mmol/L, vitamin C < 0.5 g/L, Hb < 10 g/L, ammonium ion < 25 mmol/L, potassium ion < 10 mmol/L, magnesium ion < 5 mmol/L, calcium ion < 5 mmol/L, zinc ion < 50 mmol/L and copper ion < 50 mmol/L.Enzymic method meets clin. requirements with good repeatability, accuracy, stability, linearity and high anti-interference.